Manuals and User Guides for Fluke 87. We have 3 Fluke 87 manuals available for free PDF download: User Manual, Service Manual Fluke 87 User Manual (96 pages) Fluke 87 Manuals The Fluke Model 87 and Model 89 Series IV True RMS Multimeters (hereafter referred to as the “meter”) comply with:. EN61010.1:1993.
- Location
- HB, CA (19 Hrs. 22 Min. from Winged Horses)
ok... i've got a fluke 87 that was giving me display fits, missing characters,
segments of numbers, etc.
i found a fix online for it, i believe it will work for any 80 series fluke DMM.
open up the case... 3 screws on the back.
separate case halves, remove battery.
remove single screw holding inner housings on circuit board...
carefully remove bezel from around LCD display
remove the back inner housing, separating the power connection from the LCD
flex the four clips on the front inner housing, separating it from the circuit board
remove the lcd from the top inner housing.....
now the fix... lightly run a pencil eraser across the flexible connectors top and
bottom of the LCD. run pencil eraser across the connection tabs on the PCB,
two strips of them....
snap everything back together, with a new battery.
voila! works like new again. check it against a trusted meter to make sure
you didn't bump any of the calibration pots....
takes about 10 minutes to do
segments of numbers, etc.
i found a fix online for it, i believe it will work for any 80 series fluke DMM.
open up the case... 3 screws on the back.
separate case halves, remove battery.
remove single screw holding inner housings on circuit board...
carefully remove bezel from around LCD display
remove the back inner housing, separating the power connection from the LCD
flex the four clips on the front inner housing, separating it from the circuit board
remove the lcd from the top inner housing.....
now the fix... lightly run a pencil eraser across the flexible connectors top and
bottom of the LCD. run pencil eraser across the connection tabs on the PCB,
two strips of them....
snap everything back together, with a new battery.
voila! works like new again. check it against a trusted meter to make sure
you didn't bump any of the calibration pots....
takes about 10 minutes to do
Fluke 87-III Industrial Multimeter
Please select an option below
Please select a quantity
Product Details
The Fluke 87-III Industrial Multimeter is a True RMS responding device that can measure AC/DC voltage to 1000V and AC/DC current to 10A. AC crest factor can be up to 3 at full scale, 6 at half scale.
Fluke 87-III Industrial Multimeter features:
- Volts, ohms, amps, continuity and diode test
- 250 microsecond Peak Hold
- 4-digit high resolution mode (19,999 counts)
- Backlit display
- Min/Max/Avg recording
- Frequency and duty cycle measurements
- Conductance measurements
- Capacitance measurements
- Relative (change) readings
- Sleep mode
- 10 Megohm input impedance
Digital Manual Request
Fluke 87-III Industrial Multimeter
Manufacturer: Fluke
Model: 87-III
Fluke 337 Clamp Meter
Fluke 87 Iii Multimeter Manual
The Fluke 337 Clamp Meter is a True RMS 1000A clamp meter wi...
Fluke 87 Industrial Multimeter
Fluke 87 Iii Manual
The Fluke 87 Industrial Multimeter with True RMS, allows you...